Effectiveness of Initial Fixation of a Grasping Pin for Proximal Femoral Fractures

https://showa.repo.nii.ac.jp/records/652
https://showa.repo.nii.ac.jp/records/652
991848d7-9311-4a53-9836-61192321b5ff
名前 / ファイル ライセンス アクション
S25_143.pdf S25_143.pdf (730.4 kB)
Item type 学内発行雑誌 / Departmental Bulletin Paper(1)
公開日 2016-11-18
タイトル
タイトル Effectiveness of Initial Fixation of a Grasping Pin for Proximal Femoral Fractures
言語
言語 eng
資源タイプ
資源タイプ departmental bulletin paper
著者 KAJI, Yasutaka

× KAJI, Yasutaka

KAJI, Yasutaka

Search repository
NAKAMURA, Masanori

× NAKAMURA, Masanori

NAKAMURA, Masanori

Search repository
MIYAOKA, Hideyo

× MIYAOKA, Hideyo

MIYAOKA, Hideyo

Search repository
NAKAMURA, Yusuke

× NAKAMURA, Yusuke

NAKAMURA, Yusuke

Search repository
INAGAKI, Katsunori

× INAGAKI, Katsunori

INAGAKI, Katsunori

Search repository
OTSUKA, Naruhito

× OTSUKA, Naruhito

OTSUKA, Naruhito

Search repository
OSHII, Hiroya

× OSHII, Hiroya

OSHII, Hiroya

Search repository
TANABE, Yuji

× TANABE, Yuji

TANABE, Yuji

Search repository
書誌情報 The Showa University journal of medical sciences

巻 25, 号 2, p. 143-154, 発行日 2013-06
抄録
内容記述タイプ Abstract
内容記述 We developed a grasping pin with a hook for osteosynthesis of proximal femoral fractures and compared its performance with that of a lag screw. Cyclic compressive tests were performed to simulate cut-outs, and quasi-static torsion, tests were conducted to simulate rotational displacement in polyurethane model bones and femoral heads collected after hip replacement surgery, and cadaveric femoral heads. In the polyurethane model bones and femoral head collected after hip replacement surgery, implant displacement was increased in the cut-out simulation test in both the grasping pin group and lag screw group, deformation was less in the grasping pin group than in the lag screw group. In polyurethane bones and cadaveric bones, the grasping pins showed higher rotational resistance compared with the lag screws in the quasi-static torsion test because of the high compression force generated during implantation. In contrast, in the collected femoral head after hip replacement surgery model, the lag screws destroyed bone tissue, the lag screw group exhibited a higher rotational resistance and a lower risk of rotational displacement than the grasping pin model. The depth of cadaveric femoral heads was 60mm compared with 30mm for femoral heads obtained after surgery; therefore, the pins could be completely inserted up to the octagonal portion in the cadaveric bones, resulting in higher rotational resistance.
DOI
関連識別子 10.15369/sujms.25.143
出版者
出版者 Showa University Society
ISSN
収録物識別子 0915-6380
著者版フラグ
出版タイプ VoR
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 12:39:34.907875
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3