{"_buckets": {"deposit": "13057093-f97c-4e46-814e-8487e8ca2501"}, "_deposit": {"created_by": 1, "id": "69", "owners": [1], "pid": {"revision_id": 0, "type": "depid", "value": "69"}, "status": "published"}, "_oai": {"id": "oai:showa.repo.nii.ac.jp:00000069", "sets": ["18"]}, "author_link": ["307", "309", "306", "301", "303", "302", "305", "308", "304"], "item_10002_biblio_info_7": {"attribute_name": "書誌情報", "attribute_value_mlt": [{"bibliographicIssueDates": {"bibliographicIssueDate": "2017-12", "bibliographicIssueDateType": "Issued"}, "bibliographicIssueNumber": "6", "bibliographicPageEnd": "732", "bibliographicPageStart": "725", "bibliographicVolumeNumber": "77", "bibliographic_titles": [{"bibliographic_title": "昭和学士会雑誌"}]}]}, "item_10002_description_6": {"attribute_name": "抄録", "attribute_value_mlt": [{"subitem_description": "従来の人工橈骨頭は関節面の形状が円形で非解剖学的であるため,接触する上腕骨小頭の変形や前腕部痛を来すことが臨床上の問題点であった.近年,解剖学的形状を再現した楕円形の人工橈骨頭が開発され,上腕骨小頭への接触面積を大きくし接触圧を低下させる効果が期待できるようになった.しかし,当インプラントは実際の挿入手技で起こりえる橈骨軸に対する設置角度の誤差により,その効果を発揮できずに悪影響を来す可能性がある.本研究の目的は,解剖学的形状の人工橈骨頭を回旋設置した場合に,接触面積,接触圧にどのように影響するかを調査することである.ホルマリン固定された学生実習用解剖屍体の上腕骨と橈骨を用い,5つの異なる設置角度(回旋0°,前方回旋15°,30°,後方回旋15°,30°)で接触面積と接触圧を100Nの軸圧をかけて測定した.回旋誤差0°の正常設置が最も接触面積が大きく平均接触圧は低かった.回旋設置角度の誤差が大きくなると接触面積が低下し,平均接触圧が高くなる傾向が得られたが有意な差はなかった.本研究では解剖学的形状の人工橈骨頭は角度誤差が大きくなると接触面積がより低下し,接触圧が高くなる傾向があったが,その程度は顕著でなかった.当インプラントは,極力正確に設置することが望ましいが,今回の実験結果からは,回旋角度誤差が生じたとしても接触面積・接触圧が顕著に不良にはならなかった.", "subitem_description_type": "Abstract"}]}, "item_10002_publisher_8": {"attribute_name": "出版者", "attribute_value_mlt": [{"subitem_publisher": "昭和大学学士会"}]}, "item_10002_relation_14": {"attribute_name": "DOI", "attribute_value_mlt": [{"subitem_relation_type": "isIdenticalTo", "subitem_relation_type_id": {"subitem_relation_type_id_text": "10.14930/jshowaunivsoc.77.725", "subitem_relation_type_select": "DOI"}}]}, "item_10002_source_id_11": {"attribute_name": "NCID", "attribute_value_mlt": [{"subitem_source_identifier": "AA12635661", "subitem_source_identifier_type": "NCID"}]}, "item_10002_source_id_9": {"attribute_name": "ISSN", "attribute_value_mlt": [{"subitem_source_identifier": "2187-719X", "subitem_source_identifier_type": "ISSN"}]}, "item_10002_text_26": {"attribute_name": "アイテムタイプ", "attribute_value_mlt": [{"subitem_text_value": "学内発行雑誌 / Departmental Bulletin Paper"}]}, "item_10002_text_27": {"attribute_name": "POS_INDEX", "attribute_value_mlt": [{"subitem_text_value": "学内発行雑誌/昭和学士会雑誌/77巻(2017)/6号"}]}, "item_10002_version_type_20": {"attribute_name": "著者版フラグ", "attribute_value_mlt": [{"subitem_version_resource": "http://purl.org/coar/version/c_970fb48d4fbd8a85", "subitem_version_type": "VoR"}]}, "item_creator": {"attribute_name": "著者", "attribute_type": "creator", "attribute_value_mlt": [{"creatorNames": [{"creatorName": "安川, 泰樹"}], "nameIdentifiers": [{"nameIdentifier": "301", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "富田, 一誠"}], "nameIdentifiers": [{"nameIdentifier": "302", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "川崎, 恵吉"}], "nameIdentifiers": [{"nameIdentifier": "303", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "根本, 哲也"}], "nameIdentifiers": [{"nameIdentifier": "304", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "金澤, 臣晃"}], "nameIdentifiers": [{"nameIdentifier": "305", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "稲垣, 克記"}], "nameIdentifiers": [{"nameIdentifier": "306", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "山田, 克俊"}], "nameIdentifiers": [{"nameIdentifier": "307", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "中原, 大輔"}], "nameIdentifiers": [{"nameIdentifier": "308", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "田邊, 裕治"}], "nameIdentifiers": [{"nameIdentifier": "309", "nameIdentifierScheme": "WEKO"}]}]}, "item_files": {"attribute_name": "ファイル情報", "attribute_type": "file", "attribute_value_mlt": [{"accessrole": "open_date", "date": [{"dateType": "Available", "dateValue": "2019-07-26"}], "displaytype": "detail", "download_preview_message": "", "file_order": 0, "filename": "77_725.pdf", "filesize": [{"value": "825.9 kB"}], "format": "application/pdf", "future_date_message": "", "is_thumbnail": false, "licensetype": "license_free", "mimetype": "application/pdf", "size": 825900.0, "url": {"label": "77_725.pdf", "url": "https://showa.repo.nii.ac.jp/record/69/files/77_725.pdf"}, "version_id": "949f69e8-79ae-4f13-a6cf-7022f00ea595"}]}, "item_language": {"attribute_name": "言語", "attribute_value_mlt": [{"subitem_language": "jpn"}]}, "item_resource_type": {"attribute_name": "資源タイプ", "attribute_value_mlt": [{"resourcetype": "departmental bulletin paper", "resourceuri": "http://purl.org/coar/resource_type/c_6501"}]}, "item_title": "解剖学的形状人工橈骨頭の回旋設置角度の違いによる接触面積・圧の影響", "item_titles": {"attribute_name": "タイトル", "attribute_value_mlt": [{"subitem_title": "解剖学的形状人工橈骨頭の回旋設置角度の違いによる接触面積・圧の影響"}]}, "item_type_id": "10002", "owner": "1", "path": ["18"], "permalink_uri": "https://showa.repo.nii.ac.jp/records/69", "pubdate": {"attribute_name": "公開日", "attribute_value": "2018-06-18"}, "publish_date": "2018-06-18", "publish_status": "0", "recid": "69", "relation": {}, "relation_version_is_last": true, "title": ["解剖学的形状人工橈骨頭の回旋設置角度の違いによる接触面積・圧の影響"], "weko_shared_id": -1}
解剖学的形状人工橈骨頭の回旋設置角度の違いによる接触面積・圧の影響
https://showa.repo.nii.ac.jp/records/69
https://showa.repo.nii.ac.jp/records/69d3f388e8-73e6-4146-b99e-7614f790f155
名前 / ファイル | ライセンス | アクション |
---|---|---|
77_725.pdf (825.9 kB)
|
|
Item type | 学内発行雑誌 / Departmental Bulletin Paper(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2018-06-18 | |||||
タイトル | ||||||
タイトル | 解剖学的形状人工橈骨頭の回旋設置角度の違いによる接触面積・圧の影響 | |||||
言語 | ||||||
言語 | jpn | |||||
資源タイプ | ||||||
資源タイプ | departmental bulletin paper | |||||
著者 |
安川, 泰樹
× 安川, 泰樹× 富田, 一誠× 川崎, 恵吉× 根本, 哲也× 金澤, 臣晃× 稲垣, 克記× 山田, 克俊× 中原, 大輔× 田邊, 裕治 |
|||||
書誌情報 |
昭和学士会雑誌 巻 77, 号 6, p. 725-732, 発行日 2017-12 |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | 従来の人工橈骨頭は関節面の形状が円形で非解剖学的であるため,接触する上腕骨小頭の変形や前腕部痛を来すことが臨床上の問題点であった.近年,解剖学的形状を再現した楕円形の人工橈骨頭が開発され,上腕骨小頭への接触面積を大きくし接触圧を低下させる効果が期待できるようになった.しかし,当インプラントは実際の挿入手技で起こりえる橈骨軸に対する設置角度の誤差により,その効果を発揮できずに悪影響を来す可能性がある.本研究の目的は,解剖学的形状の人工橈骨頭を回旋設置した場合に,接触面積,接触圧にどのように影響するかを調査することである.ホルマリン固定された学生実習用解剖屍体の上腕骨と橈骨を用い,5つの異なる設置角度(回旋0°,前方回旋15°,30°,後方回旋15°,30°)で接触面積と接触圧を100Nの軸圧をかけて測定した.回旋誤差0°の正常設置が最も接触面積が大きく平均接触圧は低かった.回旋設置角度の誤差が大きくなると接触面積が低下し,平均接触圧が高くなる傾向が得られたが有意な差はなかった.本研究では解剖学的形状の人工橈骨頭は角度誤差が大きくなると接触面積がより低下し,接触圧が高くなる傾向があったが,その程度は顕著でなかった.当インプラントは,極力正確に設置することが望ましいが,今回の実験結果からは,回旋角度誤差が生じたとしても接触面積・接触圧が顕著に不良にはならなかった. | |||||
DOI | ||||||
関連識別子 | 10.14930/jshowaunivsoc.77.725 | |||||
出版者 | ||||||
出版者 | 昭和大学学士会 | |||||
ISSN | ||||||
収録物識別子 | 2187-719X | |||||
著者版フラグ | ||||||
出版タイプ | VoR |