Evaluation of Deep Venous Thrombosis Using Dual-energy CT
https://showa.repo.nii.ac.jp/records/487
https://showa.repo.nii.ac.jp/records/487291edd2b-50cc-46de-98f7-ffcb63761a2f
名前 / ファイル | ライセンス | アクション |
---|---|---|
S30_107.pdf (239.7 kB)
|
|
Item type | 学内発行雑誌 / Departmental Bulletin Paper(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2018-08-23 | |||||
タイトル | ||||||
タイトル | Evaluation of Deep Venous Thrombosis Using Dual-energy CT | |||||
言語 | en | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | departmental bulletin paper | |||||
著者 |
YAMANA, Keita
× YAMANA, Keita× OHGIYA, Yoshimitsu× TASHIRO, Yuki× KURODA, Hanako× TOYOFUKA, Kosuke× GOKAN, Takehiko |
|||||
書誌情報 |
The Showa University journal of medical sciences 巻 30, 号 1, p. 107-112, 発行日 2018-03 |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | Deep vein thrombosis (DVT) was evaluated using virtual monochromatic imaging (VMI) with dual-energy analysis. We used data from 29 patients who were imaged at Showa University Hospital using the pulmonary embolism (PE)/DVT protocol and evaluated based on the diagnostic utility of VMI with dual-energy analysis. By VMI, we compared the computed tomography (CT) values of the femoral veins on both sides and the surrounding muscle tissues at 40keV and conventional imaging; two radiologists performed the visual evaluation in three stages. We also evaluated the CT values of thrombi for cases with DVT. We conducted a retrospective cohort study on 29 patients (18 males and 11 females) with a mean age of 66.8 (range: 31?86) years. DVT was confirmed in four of the subjects (13.7%). Visual evaluation confirmed a significant difference between the CT values at 40keV and with conventional imaging (2.76 vs. 1.81;?P<0.05). CT values in the femoral vein were 274.67 (range: 186.35?63.31) Hounsfield units (HU) at 40keV and 109.46 (range: 74.54?155.66) HU with conventional imaging (P<0.05). Contrast-to-noise ratios (CNRs) [(femoral vein CT value?adductor muscle CT value)/adductor muscle standard deviation (SD)] were 11.77 (range: 3.93?26.33) HU at 40keV and 9.87 (range: 1.28?27.56) HU with conventional imaging (P<0.05). The thrombus/femoral vein ratio (CT value) was 0.34 at 40keV and 0.59 with conventional imaging, while the CNR was 17.52 at 40keV and 4.32 with conventional imaging; both differences were significant. Low-voltage contrast CT is beneficial for enhancing images of veins and it may also be very useful for detecting DVT. | |||||
DOI | ||||||
関連タイプ | isIdenticalTo | |||||
識別子タイプ | DOI | |||||
関連識別子 | 10.15369/sujms.30.107 | |||||
出版者 | ||||||
出版者 | Showa University Society | |||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 0915-6380 | |||||
出版タイプ | ||||||
出版タイプ | VoR | |||||
出版タイプResource | http://purl.org/coar/version/c_970fb48d4fbd8a85 |